GENERATOR |
XRP-320/4500/2 |
|
Voltage range |
15 - 320 kV |
|
Adjustment increments (minimum step) |
0.2 kV/step |
|
Accuracy |
± 1% of output value ± 0.2 kV |
|
Reproducibility (at constant temperature) |
± 0.01% of maximum kV-value |
|
High voltage ripple
(with 10 m high voltage cable) |
10 V/mA, min. 40 V |
|
Temperature induced drift |
80 ppm/°C based on output value |
|
Current range |
0 - 22.5 mA |
|
Adjustment standard range |
in 0.05 mA steps from 0.5 mA to max. value |
|
High resolution range (recommended) |
in 0.01 mA steps from 0.0 mA to max. value |
|
Accuracy (at constant temperature) |
± 0.2% of output value ± 0.01 mA |
|
Reproducibility (at constant temperature) |
± 2 µA |
|
Temperature induced drift |
50 ppm/°C of output value |
|
Maximum power |
4500 W |
|
High voltage connector type |
R24 |
|
|
|
CONTROL UNIT XRG |
XRG |
|
Dimensions (W x H x D) |
483 mm x 133 mm x 300 mm |
|
Weight |
12.5 kg |
|
|
|
POWER SUPPLY - XRP POWER |
XRP-POWER-4500 |
|
Dimensions (W x H x D) |
340 mm x 350 mm x 628 mm |
|
Weight |
45 kg |
|
|
|
HIGH VOLTAGE TANK (OIL INSULATED) |
XRP-T160K / XRP-T320A |
|
Dimensions (W x H x D) |
375 mm x 335 mm x 625 mm |
|
Weight |
80 kg |
|
|
|
SAFETY FEATURES |
|
|
Two separate, independently monitored safety circuits (fail-safe, 24 V) |
|
Continuous system-function monitoring with fully automated system shut-down and failure indication |
|
Automated filament post-heating high voltage capacitor discharge upon termination |
|
“X-ray on” warning lamp monitoring provided (fail-safe) |
|
External safety circuit for automated systems (24 V) |
|
High voltage primary circuit interruption for CDRH switch (US standard) |
|
Monitored door lock feature |
|
External coolant monitoring |